5.  MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS 
 
Page 109 
© 2007 DH Instruments, a Fluke Company 
pressures.  Absolute pressures are be defined by adding atmospheric pressure measured 
by a barometer to gauge pressure defined by the piston gauge.  If, after calibration, the RPM4 
will not be used in absolute measurement mode (as is often the case for higher pressure 
Q-RPTs) it may be calibrated in gauge mode using a gauge pressure standard. 
 
  Q-RPTs calibrated in gauge measurement mode by applying gauge reference pressure values 
should be used in gauge mode only (see Sections 3.3.3,   PRINCIPLE, 5.3.5). 
 
•  Able to supply the recommended sequence of pressure points in the range to be 
calibrated:  See Section 5.3.4 for information on the recommended calibration point 
sequence. 
5.3.3 
SET-UP AND PREPARATION 
To set-up and prepare the RPM4 for calibration of its Q-RPTs: 
 
 
  THE RPM4 SHOULD NOT BE REMOVED FROM THE HPMS BRACKET FOR CALIBRATION.  THE 
RPM4/HPMS SHOULD ALWAYS BE KEPT TOGETHER AS AN INTEGRATED UNIT UNLESS 
DISASSEMBLY IS REQUIRED FOR MAINTENANCE OR REPAIR. 
 
  Set the RPM4/HPMS on a stable surface near the calibration standard at a height as 
close as possible to the calibration standard’s reference height.  Consider the high 
pressure gas connection that needs to be made to the TEST port on the rear of the HPMS 
and access to the front panel display, keypad and Lo Q-RPT isolation valve. 
  Connect the calibration standard output to the HPMS rear TEST  port.  The TEST port 
connection is DH500 F (DH500 F:  gland and collar type fitting for coned and left hand 
threaded 1/4 in. OD tube, equivalent to AE F250C, HIP HF4, etc.).  The HPMS TEST port 
connects to BOTH of the RPM4 A70M/A20M-AF quartz reference pressure transducers 
(see Section 3.2.5). 
 
  Take care to avoid overpressure of the Lo Q-RPT (A20M with maximum pressure of 3 000 
psi).    The  HPMS  rear  TEST  port  connects  to  both  the  A20M  (3 000 psi)  and  the  A70M 
(10 000 psi) Q-RPTs.  Use the Lo RPT Isolation Valve on the HPMS front panel to isolate the Lo 
RPT  when  applying  pressure  to  the  Hi  RPT  (see  Section  3.2.5).    If  RPM4  is  removed  from  the 
HPMS, note that the RPM4 has separate Lo and Hi TEST(+) ports, one for each Q-RPT.  Exposing 
the  Lo  RPT  to  pressure  greater  than  3 600 psi  (24 MPa)  may  damage  it  beyond  repair.    The 
product warranty does not cover Q-RPT damage due to overpressure.