Fluke A20M-AF ユーザーズマニュアル

ページ / 148
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 
© 2007 DH Instruments, a Fluke Company 
Page 2 
1.2 
SPECIFICATIONS 
1.2.1 
GENERAL SPECIFICATIONS 
Power Requirements: 
85 to 264 VAC, 50/60 Hz, 25 VA max consumption 
and 12VDC, 1.2 A 
Operating Temperature Range: 
18 to 28 °C 
Storage Temperature Range: 
- 20 to 70 °C 
Weight: 
11 kg (22 lb) approx. 
Dimensions: 
23.5 cm H x 29.4 cm W x 37.5 cm D (9.3 in. x 11.6 in. x 14.8 in.) 
Communication Ports 
Fuses: 
RS232 (COM1, COM2), IEEE-488.2 
1 A, 250 VAC fuse, 5 x 20 mm, time lag type fuse 
Internal power supply fuse not replaceable by operator: 2.5A, 250 VAC 
Pressure Range: 
0 to 10 000 psi (70 MPa) in gauge mode 
Atmosphere to 10 000 psi (70 MPa) in absolute mode 
Operating Medium: 
Any clean, dry, non-corrosive gas 
Pressure Connections: 
DH500 female 
 
  DH500 is a gland and collar type fitting for 6mm (1/4 in.) coned and left 
hand threaded tubes equivalent to AE F250C, HIP HF4, etc. 
 
Pressure Limits: 
Maximum Working
 
Pressure: 10 000 psi (70 MPa) 
Maximum Pressure Without Damage: 12 000 psi (83 MPa) 
When low Q-RPT is connected to pressure: 
Maximum Working
 
Pressure: 3 000 psi (20 MPa) 
Maximum Pressure Without Damage: 3 600 psi (24 MPa) 
 
PGC-10000-AF SYSTEM
 
Temperature: 
Operating: 
18 to 28 
°C 
 
Storage: 
-20 to 70 
°C 
Relative Humidity: 
Operating: 
15 to 70%RH (non-condensing) 
 
Storage: 
10 to 90%RH (non-condensing)