RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 
© 2007 DH Instruments, a Fluke Company 
Page 20 
The ranges available on the RPM4 are accessed using [RANGE] (see Section 3.3.1) and/or 
created using [AutoRange] (see Section 3.3.4). 
Most settings made in an RPM4 range, such as unit of measure, measurement mode, display 
resolution, and stability setting are specific to the range.  Settings selected while one range is 
active apply to that range and not to other ranges.  The range specific settings are stored with 
the range and recalled whenever the range is made active.  This makes setting up ranges a 
convenient way to store and recall frequently used operating configurations.  See Table 2 for a listing 
of RPM4 adjustments and settings and whether they are range, Q-RPT or system specific. 
Table 2.  Settings and what they are specific to 
(range, measurement mode, Q-RPT, system) 
SETTING 
PURPOSE 
SPECIFIC TO 
SECTION 
[UNIT] 
Set pressure unit of measure 
Range 
3.3.2 
[MODE] 
Set pressure measurement mode 
(absolute, gauge) 
Range 
3.3.3 
[DISPLAY] 
Set bottom line display function 
System 
3.3.6 
[HEAD] 
Set fluid head correction height, fluid, unit 
of measure 
System 
3.3.7 
[AutoZ] 
Run AutoZ 
Q-RPT and measurement mode 
3.3.9 
Resolution 
Set pressure display resolution 
Range 
3.4.2 
Stability Set 
Ready/Not Ready stability test. 
Range 
3.4.3 
Upper Limit (UL) 
Set upper limit alarm 
Range and measurement mode 
3.4.4 
AutoZ 
AutoZ, ON/OFF, set and view values 
Q-RPT and measurement mode 
3.5.1 
Screen Saver, 
Sound, Time, 
ID, Level 
Set system user preferences 
System 
3.5.5 
Cal 
Various Q-RPT and barometer calibration 
functions, including turning off absolute and 
negative gauge modes 
Q-RPT or barometer 
3.5.8 
3.2.4 
AUTOMATED TEST AND CALIBRATION SEQUENCES 
The RPM4 AutoTest function supports “Quick” and “File” automated calibration sequences.  
These automatically AutoRange the RPM4, setting its resolution, stability limit and upper limit 
based on the characteristics of the device under test.  They also prompt the user through the 
increments of the calibration sequence and log calibration data.  The AutoTest function 
should be used for most common calibration tasks, especially calibration of analog pressure 
gauges (see Section 3.3.10). 
3.2.5 
HPMS (HIGH PRESSURE MOUNTING SYSTEM) 
The HPMS (high pressure mounting system) holds the RPM4 at a convenient viewing angle 
and includes a valve and visual indicators to isolate and protect the RPM4’s 2 000 psi (A20M) 
Q-RPT when using the 10 000 psi (A70M) Q-RPT. 
See Figure 10 for an HPMS schematic. 
  OPERATION 
Numerical references in this Section refer to Figure 9. 
In normal use, the operator’s only interaction with the HPMS is to operate the isolation valve 
knob to connect and disconnect the Lo Q-RPT from the TEST port.  The valve’s operation is 
prompted by the Valve Position LEDs (2, 4).