Fluke A20M-AF ユーザーズマニュアル

ページ / 148
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL 
© 2007 DH Instruments, a Fluke Company 
Page 52 
1. 
Full scale of the AutoTest with pressure unit of 
measure and measurement mode. 
2. 
Number of test points executed/total number 
of points in the AutoTest. 
3.  Date and time that the last point in the 
AutoTest was executed. 
       
                        
 
6000 psig        6/6 
20050922 10:51:01 am 
                 
 
 
Press  [+/-] or [
] to scroll through the headers of the other AutoTest data files 
available.  RPM4 stores up to ten AutoTest data files in order of execution with a 
new data file pushing all the files back and deleting the oldest one.  When the 
header of the AutoTest data file you would like to view appears, press [ENT] to 
view the first point of the identified AutoTest data file. 
The display is: 
1. 
RPM4 reading logged for the test point. 
2. 
Head, AutoZ and Q-RPT indicator (same as in 
the RPM4 main run screen) (see Section 
3.1.1). 
3.  Indication that this is an AutoTest data file 
view screen. 
4. 
<FA>  indicating that the AutoTest was a File 
AutoTest or <QA> indicating a Quick 
AutoTest.  This is preceded by <OT>  if the 
test point was out of tolerance (definition of 
out of tolerance is Set point – RPM4 reading > 
DUT tolerance). 
5. 
Number of this test point over total number of 
points in the AutoTest. 
       
                      
 
0.00 psig      hzHi 
View data    FA 1/10 
       
                
     
 
 
Press [ENT] to move to the next data point.  Press [
] or [] to scroll backward 
and forward through the test points.  The data file header is between the first 
point and the last point. 
Press  [ESC] to exit the data file.  Confirmation is required to exit the data view 
function. 
 
  Up to ten data files are buffered, ten at a time, in RPM4 memory.  When a new 
AutoTest run is completed, its data file goes to front of the queue and the oldest 
data file is deleted. 
 
3.4.5.2 
FILE 
 PURPOSE 
To view and edit File AutoTest definition files. 
 PRINCIPLE 
RPM4 supports AutoTests (see Section 3.3.10, 
 PRINCIPLE).  There are two 
types of AutoTests: 
•  Quick AutoTests in which the DUT and test parameters are entered “on the 
fly” during the initialization of the test. 
•  File AutoTests in which the DUT and test parameters are stored in a file and 
retrieved when the file is selected to run an AutoTest. 
File AutoTest parameters are stored in definition files.  [SETUP]<5AutoTest>
<2file>  allows File AutoTest definition files to be viewed and edited.  Up to 40 
File AutoTest definition files with up to 21 test points in each file can be created.