4. REMOTE OPERATION 
 
Page 77 
© 2007 DH Instruments, a Fluke Company 
4.4 
COMMANDS 
4.4.1 
PROGRAMMING MESSAGES 
Some commands accept an optional suffix. This suffix can be used to specify the explicit Q-RPT  
to address. In most cases, each Q-RPT in the RPM4 operates as an independent measurement 
device. If the suffix is not specified, then the “Active” Q-RPT is assumed. 
 
 
The “RANGE” or “ARANGE” command can be used to select the “Active” Q-RPT. See the 
program message descriptions for suffix use with a specific command.  
Table 14.  Program message list 
COMMAND 
n’  indicates an 
optional suffix 
DESCRIPTION 
Send a command string out of the RPM4 COM2 port. 
ABORT 
Stop pending operations 
ARANGEn 
Read or set a new AutoRange to use. 
ATM 
Read the current atmospheric pressure (on-board barometer). 
AUTOZEROn 
Read or set the status of the AutoZ automatic zeroing function. 
AUTOZEROn=RUN 
Run AutoZero. 
CALAMB 
Read or set the on-board barometer calibration 
COM1 
Read or set the configuration of the COM1 port. 
COM2 
Read or set the configuration of the COM2 port. 
CONTn 
Enable continuous measurements 
DATE 
Read or set the current date. 
ERR 
Read the last error message. 
GPIB 
Read or set the GPIB interface address. 
HEAD 
Read or set the fluid head settings. 
ID 
Read or set the RPM4 alphanumeric asset ID tag. 
L2 
Selects “classic” program message format 
L3 
Selects “enhanced” program message format 
LOCAL 
Return control to the RPM4 front panel. 
MEM 
Read the power-up memory test status. 
MMODEn 
Read or change the active measurement mode. 
MSGFMT 
Read or set the type of program message format to use. 
PCALn 
Read or set the user Lo Q-RPT calibration information. 
PCAL:XX 
Read or set the user Lo Q-RPT calibration information (old command). 
PRn 
Read the next RPM4 pressure. 
PRRn 
Read the next RPM4 pressure, rate, and ATM. 
QPRRn 
Read the last RPM4 pressure, rate and ATM. 
RANGE 
Select a Q-RPT to be “active” in it’s default full scale range 
RATEn 
Read the next available rate of change of pressure. 
READRATEn 
Read or set the Q-RPT measurement read rate and mode 
READYCKn 
Read or set a flag that is cleared by a Not Ready condition. 
REMOTE 
Enable remote local lockout operation. 
RESn 
Read or set the pressure display resolution for the current transducer and range. 
RESET 
Reset the RPM4 to default user parameters. 
RPTn 
Read the available Q-RPT data. 
SCRSAV 
Read or set the front panel screen saver period. 
SN 
Read the serial number of the RPM4. 
SRn 
Read the next available pressure status (Ready/Not Ready). 
SSn 
Read or set the stability required for a Ready condition. 
SS%n 
Read or set the stability required for a Ready condition (% span/ s). 
TIME 
Read or set the current time of day. 
UCOEFn 
Convert a pressure in Pascal to pressure in the current units. 
UDU 
Read or set the user defined pressure unit. 
ULn 
Read or set the upper limit for the current range. 
UNITn 
Read or set the pressure unit of measure for the current range. 
VER 
Read the RPM4 software version. 
ZOFFSETn 
Read or set the AutoZero  Z
offset
 for the specified Q-RPT. 
ZOFFSET:XX 
Read or set the AutoZ Z
offset
 for the specified Q-RPT (old command).