Siemens LU01 ユーザーズマニュアル

ページ / 114
Page 76
SITRANS LU01 – INSTRUCTION MANUAL
7ML19985BE02
Pa
ra
m
e
te
rs
Select the TVT type which gives the highest confidence (P805) under all level conditions. 
Use this parameter with caution, and do not use TVT Slopes with the bF or bLF Algorithm 
(P820). 
P831 TVT Shaper
Turns the TVT Shaper ON or OFF.
Turn the TVT Shaper ON before using P832 and afterwards. Turn the TVT Shaper ON and 
OFF while monitoring the effect to pick up the true echo. 
P832 TVT Shaper Adjust
Allows manual adjustment of the TVT curve.
Use this feature to bias the shape of the TVT curve to avoid selecting false echoes from 
fixed objects.
Adjust this parameter while viewing the echo profile with Dolphin Plus. Refer to the 
Dolphin Plus online help for details. If Dolphin Plus is not available, then an oscilloscope 
can be used. When using an oscilloscope, the Echo Lock Window display becomes the 
TVT Curve Pointer. See 
P810 Scope Displays
 on page 71 for more information.
The TVT curve is divided into 40 breakpoints, accessible by enabling the point number as 
the breakpoint index field. Each breakpoint is normalized to a value of 0, as displayed in 
the parameter value field. By changing the breakpoint value, up or down, the intensity of 
the bias applied to that breakpoint of the curve is respectively changed. By changing the 
value of adjacent breakpoints, the effective bias to the shaper can be broadened to suit 
the desired correction. In the case of multiple false echoes, shaping can be applied along 
different points of the curve. Shaping should be applied sparingly in order to avoid 
missing the true echo.
To change a breakpoint
1.
Confirm that P831, TVT shaper, is on.
2.
Go to P832.
3.
Press MODE 
 and then press DISPLAY 
 to access the Breakpoints.
4.
Press MODE 
 to select the Auxiliary function.
5.
Press 
 to scroll the breakpoint index through points 1 to 40, and move the 
curve pointer right or left respectively.
Values
0
*
Off
1
On
Related
P832 TVT Shaper Adjust
Primary Index
Breakpoint
Values
Range: -50 to 50
Preset: 0
Related
P810 Scope Displays
P831 TVT Shaper
P
P