Siemens 2000592-001 ユーザーズマニュアル

ページ / 72
Site Wiring Requirements, 
Continued 
 
DataNET Channels 
The DataNET channel connections can be either copper wire or fiber 
optic cable. The wiring method and cable type for DataNET is dependent 
upon the site requirements. For cable requirements see Table 1-1. 
The following Table should be used when evaluating or selecting cable 
for an existing or new DataNET installation. 
Table 1-1:  
Recommended DataNET 
Cable  
Cable Type 
Vendors & P/N 
Installation Notes 
Fiber Optic 
MM 62.5/125; 2 Fiber 
Conductors 
SEA P/N 1270002-002 
Industry Available from 
Multiple Vendors 
New Installation as 
an alternative to 
copper cable 
Max distance from 
Analyzer to ANG or 
between Hubs 
1520 meters (5000 
ft) 
2- Conductor Copper 
Wire 
SEA P/N 1686002-001 
Belden P/N 9182 
For 
Belden Armored Cable 
order 
SEA P/N 1680000-002 
Recommended for 
Single Channel 
Max distance from 
Analyzer to ANG or 
between Hubs 
1520 meters (5000 
ft) 
4-Conductor Copper 
Wire 
SEA P/N 1681000-003 
Belden P/N SSD1743 
Recommended for 
Dual Channel 
Max distance from 
Analyzer to ANG or 
between Hubs 
1520 meters (5000 
ft) 
 
8  
  Installation 