7.4
Recording: Duration and Type – durA, CYCLE
•
Recording without Time Limits – durA
Key Sequence: SEt
... ... durA
on
oFF
Settings:
oFF = time limiting disabled
Recording is continued until the memory is full. If the CYCLE function has been activated,
memory is overwritten in a cyclical fashion. See also drawing in chapter 7.3, page 16.
memory is overwritten in a cyclical fashion. See also drawing in chapter 7.3, page 16.
•
Time-Limited Recording – durA
Key Sequence: SEt
... ... durA
ON
dd
hh:mm
Settings:
d = number of days (adjustable from 0 to 6)
hh = number of hours, mm = number of minutes
hh = number of hours, mm = number of minutes
Recording is ended after the predetermined duration has expired.
See also drawing in chapter 7.3, page 16.
See also drawing in chapter 7.3, page 16.
•
Cyclical Recording – CYCLE
Key Sequence: SEt
rAtE ... CYCLE
OFF
on
Settings:
on = cyclical recording activated
When the memory is full, the oldest data are overwritten with new data (FIFO). See also
drawing in chapter 7.3, page 16.
drawing in chapter 7.3, page 16.
•
Record Until Memory is Full – CYCLE
Key Sequence: SEt
... ... CYCLE
on
OFF
Settings:
OFF = cyclical recording deactivated
When the memory full, data recording is ended. See also drawing in chapter 7.3, page 16.
7.5
Internal Time – timE
•
Set Current Weekday and Time (no date)
Key Sequence: SEt
... ... tiME
dd
hh:mm
Settings:
dd = current weekday
hh:mm = current time in hours and minutes
hh:mm = current time in hours and minutes
Current weekday and time are set with this function.
7.6
Select Address and Modem Option – Addr, ModEM
Key Sequence: SEt
... ... Addr
nn
ModEM
no
YES
Settings:
nn = 1 ... 15
ModEM = YES, if the adapter is
connected to the PC with a modem
ModEM = YES, if the adapter is
connected to the PC with a modem
If several memory adapters are connected to the PC, a separate address must be assigned
to each. Address number 1 should be assigned to the first adapter, address number 2 to the
second and so forth.
Note: If only one adapter has been connected, it should be set to address number 1.
to each. Address number 1 should be assigned to the first adapter, address number 2 to the
second and so forth.
Note: If only one adapter has been connected, it should be set to address number 1.