Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
1-94
 
Making Measurements
Using Bandwidth Limits to Test a Bandpass Filter
3. Substitute a thru for the device and perform a response calibration by pressing:
 
 
 
4. Reconnect your test device.
Refer to 
.
Setting Up the Bandwidth Limits
When you set up the bandwidth limits to test the bandpass filter, you will set 
• the amplitude below the peak that is used to measure the filter’s bandwidth. This 
setting is called N dB Points
• the Maximum Bandwidth value. If the measured bandwidth is greater than this value, 
the test will fail.
• the Minimum Bandwidth value. If the measured bandwidth is less than this value, the 
test will fail.
1. To access the bandwidth menu, press:
 
 
2. To set the amplitude below the peak passband amplitude that you want to measure the 
bandwidth. In this case, we are setting the bandwidth that will be measured 40 dB 
below the peak amplitude of the bandpass filter by pressing:
 
 
3. To set the minimum bandwidth for the bandwidth test, press:
 
 
4. To set the maximum bandwidth for the bandwidth test, press:
 
 
Running a Bandwidth Test
After setting up the bandwidth limits, you are ready to run the bandwidth test and check 
the test results. For this example, we will:
• Start the test.
• Display the bandwidth markers.
• Review the test results.
Cal
 CALIBRATE MENU   RESPONSE   THRU 
System
 LIMIT MENU   BANDWIDTH LIMIT 
 N DB POINTS 
40
x1
 MINIMUM BANDWIDTH
100
M/
µ
 MAXIMUM BANDWIDTH
150
M/
µ