Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
1-98
 
Making Measurements
Using Test Sequencing
Using Test Sequencing
Test sequencing allows you to automate repetitive tasks. As you make a measurement, the 
analyzer memorizes the keystrokes. Later you can repeat the entire sequence by pressing a 
single key. Because the sequence is defined with normal measurement keystrokes, you do 
not need additional programming expertise. Subroutines and limited decision-making 
increases the flexibility of test sequences. In addition, the GPIO outputs can be controlled 
in a test sequence, and the GPIO inputs can be tested in a sequence for conditional 
branching.
The test sequence function allows you to create, title, save, and execute up to six 
independent sequences internally. You can also save sequences to disk and transfer them 
from the analyzer to another analyzer or possibly to an external computer controller (so 
the sequence can be sent to another analyzer).
How to Use Test Sequencing
The following procedures, which are based on an actual measurement example, show you 
how to do the following:
• create a sequence
• title a sequence
• edit a sequence
• clear a sequence
• change a sequence title
• name files generated by a sequence
• store a sequence
• load a sequence
• purge a sequence
• print a sequence
Creating a Sequence 
1. To enter the sequence creation mode, press:
 
As shown in 
, a list of instructions appear on the analyzer display to help 
you create or edit a sequence.
Seq
 NEW SEQ/MODIFY SEQ