Making Mixer Measurements (Option 089 Only)
Measurement Considerations
Frequency offset measurements do not begin until all of the frequency offset mode 
parameters are set. These include the following:
• Start and Stop IF Frequencies
• LO frequency
• Up Converter / Down Converter
• RF > LO / RF < LO
The LO frequency for frequency offset mode must be set to the same value as the external 
LO source. The offset frequency between the analyzer source and receiver will be set to this 
value.
For a single-sideband mixer measurement, the RF source can be offset in frequency from 
the input receiver frequency, allowing for a swept RF stimulus over one frequency range 
and measurement of the IF response over another (in this case the output IF).
When frequency offset mode operation begins, the receiver is tuned to the entered IF signal 
frequencies and then offsets the source frequency required to produce the IF. Therefore, 
since it is the analyzer receiver that controls the source, it is only necessary to set the start 
and stop frequencies from the receiver.
LO Frequency Accuracy and Stability 
The analyzer source is phaselocked to its receiver through a reference loop. In the 
frequency offset mode, the mixer under test is inserted in this loop. To ensure that the 
analyzer phaselocks correctly, it is important that you use an LO source that has frequency 
accuracy better than 
±
16 MHz and residual FM < 20 kHz RMS.
Differences Between Internal and External R Channel Inputs 
Due to internal losses in the analyzer's test set, the power measured internally at the R 
channel is 16 dB lower than that of the source. To compensate for these losses, the traces 
associated with the R channel have been offset 16 dB higher. As a result, power measured 
directly at the R channel via the R CHANNEL IN port will appear to be 16 dB higher than 
its actual value. If power meter calibration is not used, this offset in power must be 
accounted for with a receiver calibration before performing measurements.
The following steps can be performed to observe this offset in power:
1. To set the power range to manual, press:
 
 
NOTE
Since the default source output power of the 8722ES is below 0 dBm, no 
reduction in source output power is required for this procedure.
Power
 PWR RANGE MAN 
0
x1