Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
4-36
 
Printing, Plotting, and Saving Measurement Results
Saving and Recalling Instrument States
Saving and Recalling Instrument States 
Places Where You Can Save 
• analyzer internal memory
• floppy disk using the analyzer's internal disk drive
• floppy disk using an external disk drive
• IBM compatible personal computer using GPIB mnemonics
What You Can Save to the Analyzer’s Internal Memory 
The number of registers that the analyzer allows you to save depends on the size of 
associated error-correction sets, and memory traces. However, the maximum number of 
registers that can be saved to internal memory is 31. Refer to the "Preset State and 
Memory Allocation" chapter of the reference guide for further information.
You can save instrument states in the analyzer internal memory, along with the following 
list of analyzer settings. The default file names are REG<01-31>.
• error-corrections on channels 1 and 2
• displayed memory trace
• print/plot definitions
• measurement setup
❏ frequency range
❏ number of points
❏ sweep time
❏ output power
❏ sweep type
❏ measurement parameter
NOTE
When the ac line power is switched off, the internal non-volatile memory is 
retained by a battery. Refer to “Specifications and Characteristics” in the 