Optimizing Measurement Results
Increasing Measurement Accuracy
Increasing Measurement Accuracy 
The following all contribute to loss of accuracy in a measurement.
Interconnecting Cables 
Cables that connect the device under test (DUT) to the analyzer are often the most 
significant contribution to random errors of your measurement. You should frequently 
perform the following steps as a precaution against errors caused by cable 
interconnections:
• Inspect for lossy cables.
• Inspect for damaged cable connectors.
• Practice good connector care techniques.
• Minimize cable position changes between error-correction and measurements.
• Inspect for cables which dramatically change magnitude or phase response when flexed. 
(This may indicate an intermittent problem.)
Improper Calibration Techniques 
Calibrations techniques performed improperly contribute to random errors to your 
measurement. You should frequently perform the following steps as a precaution against 
errors caused by calibration techniques:
• Verify the correct calibration kit definition is selected.
• Verify the correct standards have been connected.
Sweeping Too Fast for Electrically Long Devices
It is possible to sweep too fast for electrically long devices. This will result in measurement 
error. Refer to 
.
Connector Repeatability 
Connector repeatability is a source of random measurement error. Measurement 
error-corrections do not compensate for these errors. For all connectors, you should 
frequently perform the following steps as a precaution against errors caused by connector 
repeatability:
• Inspect the connectors.
• Clean the connectors.
• Gauge the connectors.
• Use correct connection techniques. (Refer to 
.)