Calibrating for Increased Measurement Accuracy
Enhanced Frequency Response Error Correction
Enhanced Frequency Response Error Correction 
The enhanced frequency response error correction removes the following errors in the 
forward direction in ET models or in both the forward and reverse directions in ES models:
• removes directivity errors of the test setup
• removes source match errors of the test setup
• removes isolation errors of the test setup (optional)
• removes frequency response of the test setup
The enhanced reflection error correction may be used to remove load match from the test 
setup when measuring bilateral devices. Enhanced reflection terms are mathematically 
derived during all enhanced response calibrations but are not applied unless initiated by 
the 
 softkey. Enhanced reflection correction is applied after the 
enhanced frequency response error correction is finished.
IMPORTANT
Use enhanced reflection error correction only on bilateral devices. A bilateral 
device has similar forward and reverse transmission characteristics. 
Examples of bilateral devices are passive devices (filters, attenuators, and 
switches). Most active devices (amplifiers) and some passive devices (isolators 
and circulators) are not bilateral. If this error correction is used for a 
non-bilateral device, errors will occur in the resulting measurement.
1. Press .
2. Select the type of measurement you want to make.
❏ If you want to make measurements in the forward direction (S
21
 S
11
), press:
 
 or on ET models: 
❏ For ES analyzers, if you want to make measurements in the reverse direction (S
12
 
S
22
), press:
 
3. Set any measurement parameters that you want for the device measurement: power, 
format, number of points, or IF bandwidth.
4. To access the measurement correction menus, press:
5. If your calibration kit is different than the kit specified under the 
 softkey, 
press:
 
 (select your type of kit) 
If your type of calibration kit is not listed in the displayed menu, refer to 
 ENH. REFL. on OFF 
Preset
Meas
 Trans: FWD S21 (B/R) 
 TRANSMISSN 
Meas
 Trans: REV S12 (A/R) 
Cal
 CAL KIT [ ] 
 CAL KIT   SELECT CAL KIT 
 RETURN