1- 19
Making Measurements
Using Display Functions
Using Memory Traces and Memory Math Functions 
The analyzer has four available memory traces, one per channel. Memory traces are totally 
channel dependent: channel 1 cannot access the channel 2 memory trace or vice versa. 
Memory traces can be saved with instrument states: one memory trace can be saved per 
channel for each saved instrument state. There are up to 31 save/recall registers available, 
so the total number of memory traces that can be present is 128 including the four active 
for the current instrument state. The memory data is stored as full precision, complex 
data. Memory traces must be displayed in order to be saved with instrument states.
Additional data can be stored onto 3.5-inch floppy disks using the front panel disk drive.
NOTE
You may not be able to store 31 instrument states if they include a large 
amount of calibration data. The calibration data contributes considerably to 
the size of the instrument state file and therefore the available memory may 
be full prior to filling all 31 registers.
Two trace math operations are implemented:
 (data/memory)
 (data
memory)
(Note that normalization is 
 not 
.) Memory traces are saved and 
recalled and trace math is done immediately after error-correction. This means that any 
data processing done after error-correction, including parameter conversion, time domain 
transformation (Option 010), scaling, etc., can be performed on the memory trace. You can 
also use trace math as a simple means of error-correction, although that is not its main 
purpose.
All data processing operations that occur after trace math, except smoothing and gating, 
are identical for the data trace and the memory trace. If smoothing or gating is on when a 
memory trace is saved, this state is maintained regardless of the data trace smoothing or 
gating status. If a memory trace is saved with gating or smoothing on, these features can 
be turned on or off in the memory-only display mode.
The actual memory for storing a memory trace is allocated only as needed. The memory 
trace is cleared on instrument preset, power on, or instrument state recall.
If sweep mode or sweep range is different between the data and memory traces, trace math 
is allowed, and no warning message is displayed. If the number of points in the two traces 
is different, the memory trace is not displayed nor rescaled. However, if the number of 
points for the data trace is changed back to the number of points in the memory, the 
memory trace can then be displayed.
If trace math or display memory is requested and no memory trace exists, the message 
CAUTION: NO VALID MEMORY TRACE
 is displayed.
To Save a Data Trace to the Display Memory 
Press 
 
 to store the current active measurement data in the 
memory of the active channel. The data trace is now also the memory trace. You can use a 
memory trace for subsequent math manipulations.
 DATA/MEM 
 DATA-MEM 
 DATA/MEM 
 DATA-MEM 
Display
 DATA
MEMORY