Calibrating for Increased Measurement Accuracy
LRM Error Correction
Perform the LRM Calibration
1. You must have a LRM calibration kit defined and saved in the USER KIT, as shown in 
NOTE
This must be done before performing the following sequence.
2. Press 
  
      
 
.
3. To measure the "LRM THRU," connect the "zero length" transmission line between the 
two test ports.
4. To make the necessary four measurements, press:
5. To measure the "LRM SHORT," connect the short to PORT 1, and press:
6. Connect the short to PORT 2, and press:
NOTE
If loads can be connected to both port 1 and port 2 simultaneously, then the 
following LRM load measurement can be performed using the 
 softkey.
7. To measure the "LRM LOAD," disconnect the short and connect the LRM load to 
PORT1.
8. Press 
 
 to access the No Loads menu. When the 
displayed trace settles, press the softkey corresponding to the load used. If a sliding load 
is used, press 
 to access the Sliding Load menu. Position the slide and press 
.
9. When all the appropriate load measurements are complete, the load data is measured 
and the 
 softkey label is underlined.
10.Connect the load to PORT 2 and press 
.
11.Repeat the previous LRM load measurement steps for PORT 2.
12.After the measurement is complete, press:
13.To measure the "ISOLATION" class, press:
❏ You could choose not to perform the isolation measurement by pressing 
 
.
Cal
 CAL KIT   SELECT CAL KIT   USER KIT   RETURN 
 RETURN 
 CALIBRATE MENU   TRL*/LRM* 2-PORT 
 LRMTHRU 
 S11 REFL LRMSHORT 
 S22 REFL LRMSHORT 
 DO BOTH FWD+REV 
 LINE/MATCH   LN/MATCH1 LOAD 
 SLIDING 
 SLIDE IS SET 
 LN/MATCH1 LOAD 
 LN/MATCH2 LOAD 
 DONE LINE/MATCH 
 ISOLATION 
 OMIT ISOLATION   DONE TRL/LRM