Agilent Technologies 20ET ユーザーズマニュアル
6-56
Calibrating for Increased Measurement Accuracy
LRM Error Correction
Perform the LRM Calibration
1. You must have a LRM calibration kit defined and saved in the USER KIT, as shown in
NOTE
This must be done before performing the following sequence.
2. Press
.
3. To measure the "LRM THRU," connect the "zero length" transmission line between the
two test ports.
4. To make the necessary four measurements, press:
5. To measure the "LRM SHORT," connect the short to PORT 1, and press:
6. Connect the short to PORT 2, and press:
NOTE
If loads can be connected to both port 1 and port 2 simultaneously, then the
following LRM load measurement can be performed using the
following LRM load measurement can be performed using the
softkey.
7. To measure the "LRM LOAD," disconnect the short and connect the LRM load to
PORT1.
8. Press
to access the No Loads menu. When the
displayed trace settles, press the softkey corresponding to the load used. If a sliding load
is used, press
is used, press
to access the Sliding Load menu. Position the slide and press
.
9. When all the appropriate load measurements are complete, the load data is measured
and the
softkey label is underlined.
10.Connect the load to PORT 2 and press
.
11.Repeat the previous LRM load measurement steps for PORT 2.
12.After the measurement is complete, press:
13.To measure the "ISOLATION" class, press:
❏ You could choose not to perform the isolation measurement by pressing
.
Cal
CAL KIT SELECT CAL KIT USER KIT RETURN
RETURN
CALIBRATE MENU TRL*/LRM* 2-PORT
LRMTHRU
S11 REFL LRMSHORT
S22 REFL LRMSHORT
DO BOTH FWD+REV
LINE/MATCH LN/MATCH1 LOAD
SLIDING
SLIDE IS SET
LN/MATCH1 LOAD
LN/MATCH2 LOAD
DONE LINE/MATCH
ISOLATION
OMIT ISOLATION DONE TRL/LRM