Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
5. If you need to calibrate with a second ECal module, connect one end of another DB25 
cable to the connector on the PC interface unit labeled "DB25 Interface to ECal 
Module B". Connect the other end of the DB25 cable to the parallel cable connector on 
the ECal module.
NOTE
Why Use a Second ECal Module?
If the frequency span of the measurement that you set up earlier exceeds the 
span of a single ECal module, you need to use another ECal module whose 
frequency range allows the rest of the measurement span to be calibrated. 
The frequency range of the ECal modules is listed in the “General 
Information” chapter of the 85097A Electronic Calibration System User's 
Guide.
6. Using an RF cable (or a microwave cable, as appropriate), connect one port of the ECal 
module to test port 1 of the analyzer. Refer to 
.
CAUTION
RF ECal modules can be damaged if you apply excessive torque to the 
connectors. Do not exceed the recommended torque indicated in the 
Electronic Calibration Module Reference Guide (part number 85091-90009). 
NOTE
It is not critical which ECal module port (Port A or Port B) is connected to the 
network analyzer test ports. The network analyzer detects where each ECal 
module port is connected and uses the appropriate module data. If Port A is 
connected to Port 1 of the analyzer, the calibration will be performed slightly 
faster.
7. Using an RF cable (or a microwave cable, as appropriate), connect the other port of the 
ECal module to test port 2 of the analyzer.
Select the ECal Options
In addition to the standard ECal method, there are two options that you may want to use 
when performing the electronic calibration. They are:
• ECal using a manual thru
Calibration using a manual thru is more accurate than calibrating using the thru 
internal to the ECal module. (A zero-length thru is created by connecting the two test 
port cables together.) The improved loss of the manual thru compared to the ECal 
module's thru, increases the accuracy of other error terms in the correction. The 
accuracy of the overall calibration is improved. Manual thrus can only be used with 
ECal modules having connectors of the opposite sex.