6-64
 
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
7. After you connect the second ECal module, press 
 to continue the 
ECal. 
8. Repeat steps 4 and 5 if you selected to calibrate using the manual thru option. 
9. Review the displayed calibration results. Refer to 
10.Save the calibration results by pressing 
.
Display the Module Information
Display information about the module by pressing 
 
 
 
.
A text window is displayed that contains the following information about the selected 
module.
1
PC interface unit port to which it is connected
6
Number of measurement points
2
Model number
7
Start (minimum) frequency
3
Serial number
8
Stop (maximum) frequency
4
Connector type
9
Suggested warmup time
5
Date of last certification
10
Warmup status (This is the remaining time 
on an internal analyzer timer. The timer 
simply counts down the recommended ECal 
module warm-up time. When the timer 
reaches zero, a message is displayed 
indicating that the ECal module is ready. The 
ECal module temperature or status is not 
read. The timer is reset when the analyzer is 
powered up or preset or when a new module 
is attached.)
 CONTINUE ECal 
Save/Recall
Cal
 ECal MENU   CONFIGURE [  ] 
 MODULE INFO