Agilent Technologies 20ET ユーザーズマニュアル
6-74
Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal (ES Analyzers Only)
Remove the Adapter
When the two sets of error correction files have been created (now referred to as
"calibration sets"), the A3 adapter may be removed.
"calibration sets"), the A3 adapter may be removed.
1. Press
to display the following menu:
2. Press
to display
and
.
also displays the internal (or external if internal not used) disk
file directory.
NOTE
In the following two steps, calibration data is recalled, not instrument states.
3. From the disk directory, choose the file associated with the port 1 error correction, then
press .
4. When this is complete, choose the file for the port 2 error correction and press
.
5. When complete, press
.
6. Enter the value of adapter A3 electrical delay by pressing
and
entering the value.
7. Select the appropriate key:
or
.
8. Press
to complete the technique for calculating the new error
coefficients and overwrite the current active calibration set in use.
This process uses up an internal memory register. The calibration in this register is not
the calibration created by adapter removal, rather it is a "scratch" calibration. You may
wish to delete the register, or re-save the new calibration in this register as shown in
the following step.
the calibration created by adapter removal, rather it is a "scratch" calibration. You may
wish to delete the register, or re-save the new calibration in this register as shown in
the following step.
9. To save the results of the new calibration set, press
.
NOTE
Adapter removal can leave a residual state in internal memory. This is not a
valid instrument state and should be deleted.
valid instrument state and should be deleted.
(This Help softkey provides a quick reference guide
to using the adapter removal technique.)
to using the adapter removal technique.)
Cal
MORE ADAPTER REMOVAL
HELP ADAPT REMOVAL
RECALL CAL SETS
ADAPTER DELAY
ADAPTER COAX
ADAPTER WAVEGUIDE
REMOVE ADAPTER
RECALL CAL SETS
RECALL CAL PORT 1
RECALL CAL PORT 2
RECALL CAL SETS
RECALL CAL PORT 1
RECALL CAL PORT 2
RETURN
ADAPTER DELAY
ADAPTER COAX
ADAPTER WAVEGUIDE
REMOVE ADAPTER
Save/Recall
SELECT DISK
INTERNAL MEMORY RETURN SAVE STATE