Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal (ES Analyzers Only)
Remove the Adapter
When the two sets of error correction files have been created (now referred to as 
"calibration sets"), the A3 adapter may be removed.
1. Press 
 
 
 to display the following menu:
2. Press 
 to display 
 and 
.
 also displays the internal (or external if internal not used) disk 
file directory.
NOTE
In the following two steps, calibration data is recalled, not instrument states.
3. From the disk directory, choose the file associated with the port 1 error correction, then 
press .
4. When this is complete, choose the file for the port 2 error correction and press 
.
5. When complete, press 
.
6. Enter the value of adapter A3 electrical delay by pressing 
 and 
entering the value.
7. Select the appropriate key: 
 or 
.
8. Press 
 to complete the technique for calculating the new error 
coefficients and overwrite the current active calibration set in use.
This process uses up an internal memory register. The calibration in this register is not 
the calibration created by adapter removal, rather it is a "scratch" calibration. You may 
wish to delete the register, or re-save the new calibration in this register as shown in 
the following step.
9. To save the results of the new calibration set, press 
 
 
 
 
.
NOTE
Adapter removal can leave a residual state in internal memory. This is not a 
valid instrument state and should be deleted.
(This Help softkey provides a quick reference guide 
to using the adapter removal technique.)
 
Cal
 MORE   ADAPTER REMOVAL 
 HELP ADAPT REMOVAL 
 RECALL CAL SETS 
 ADAPTER DELAY 
 ADAPTER COAX 
 ADAPTER WAVEGUIDE 
 REMOVE ADAPTER 
 RECALL CAL SETS 
 RECALL CAL PORT 1 
 RECALL CAL PORT 2 
 RECALL CAL SETS 
 RECALL CAL PORT 1 
 RECALL CAL PORT 2 
 RETURN 
 ADAPTER DELAY 
 ADAPTER COAX 
 ADAPTER WAVEGUIDE 
 REMOVE ADAPTER 
Save/Recall
 SELECT DISK 
 INTERNAL MEMORY   RETURN   SAVE STATE