Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
7-20
 
Operating Concepts
S-Parameters
S-Parameters 
The 
 key accesses the S-parameter menu which contains softkeys that can be used 
to select the parameters or inputs that define the type of measurement being performed.
Understanding S-Parameters 
S-parameters (scattering parameters) are a convention used to characterize the way a 
device modifies signal flow. A brief explanation of the S-parameters of a two-port device is 
provided, however, for additional details, refer to Application Notes 95-1 and 154.
S-parameters are always a ratio of two complex (magnitude and phase) quantities. 
S-parameter notation identifies these quantities using the numbering convention:
         S out in
where the first number (out) refers to the test-device port where the signal is emerging and 
the second number (in) is the test-device port where the signal is incident. For example, 
the S-parameter S
21
 identifies the measurement as the complex ratio of the signal 
emerging at the test device’s port 2 to the signal incident at the test device’s port 1.
 is a representation of the S-parameters of a two-port device, together with an 
equivalent flowgraph. In the illustration, "a" represents the signal entering the device and 
"b" represents the signal emerging. Note that a and b are not related to the A and B input 
ports on the analyzer.
Meas