Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
7- 51
Operating Concepts
Measurement Calibration
Figure 7-39 Full Two-Port Error Model Equations 
How Effective Is Accuracy Enhancement? 
In addition to the errors removed by accuracy enhancement, other systematic errors exist 
due to limitations of dynamic accuracy, test set switch repeatability, and test cable 
stability. These, combined with random errors, also contribute to total system 
measurement uncertainty. Therefore, after accuracy enhancement procedures are 
performed, residual measurement uncertainties remain.
The uncorrected performance of the analyzer is sufficient for many measurements. 
However, the next three illustrations show the improvements that can be made in 
measurement accuracy by using a more complete calibration routine. See 
, and