Operating Concepts
Calibration Routines
Enhanced Reflection Calibration
The enhanced reflection calibration is activated by selecting 
 under 
the  
menu.
The enhanced reflection calibration effectively removes load match error from the 
enhanced response calibration performed on a bilateral device. A bilateral device has an 
identical forward (S
21
) and reverse transmission (S
12
) response. Most passive devices (such 
as filters, attenuators, or switches) are bilateral. Some passive devices (circulators, 
isolators) and most active devices do not have identical forward and reverse transmission 
responses and enhanced reflection calibration will not work with these devices.
S
11
 and S
22
 One-Port Calibration 
The S
11
 and S
22
 one-port calibration procedures, activated by pressing the 
 or 
 softkey within the calibrate menu, provide directivity, source match, and 
frequency response vector error-correction for reflection measurements. These procedures 
provide high accuracy reflection measurements of one-port devices or properly terminated 
two-port devices.
Full Two-Port Calibration (ES Models Only)
The full two-port calibration, activated by pressing the 
 softkey within the 
calibrate menu, provides directivity, source match, load match, isolation, and frequency 
response vector error-correction, in both forward and reverse directions, for transmission 
and reflection measurements of two-port devices. This calibration provides the best 
magnitude and phase measurement accuracy for both transmission and reflection 
measurements of two-port devices, and requires an S-parameter test set.
In this type of calibration, both forward and reverse measurements must be made. You 
have the option of setting the ratio of the number of forward (or reverse) sweeps versus the 
number of reverse (or forward) sweeps. To access this function, press 
 
 
 and enter the number of sweeps desired.
TRL*/LRM* Two-Port Calibration 
The TRL*/LRM* two-port calibration, activated by pressing the 
 
softkey within the calibration menu, provides the ability to make calibrations using the 
TRL or LRM method. True TRL/LRM calibration is available on analyzers equipped with 
Option 400, the Four Sampler Test Set. For more information, refer to 
 ENH. REFL. ON off
 ENHANCED RESPONSE 
S11 1-PORT 
S22 1-PORT 
FULL 2-PORT 
Cal
MORE 
TESTSET SW 
TRL*/LRM* 2-PORT