Safety and Regulatory Information
Safety Considerations
General 
WARNING
To prevent electrical shock, disconnect the analyzer from mains 
before cleaning. Use a dry cloth or one slightly dampened with water 
to clean the external case parts. Do not attempt to clean internally.
WARNING
If this product is not used as specified, the protection provided by 
the equipment could be impaired. This product must be used in a 
normal condition (in which all means for protection are intact) only.
CAUTION
This product is designed for use in Installation Category II and Pollution 
Degree 2 per IEC 1010 and 664 respectively.
CAUTION
VENTILATION REQUIREMENTS: When installing the product in a cabinet, 
the convection into and out of the product must not be restricted. The ambient 
temperature (outside the cabinet) must be less than the maximum operating 
temperature of the product by 4
°
 C for every 100 watts dissipated in the 
cabinet. If the total power dissipated in the cabinet is greater that 800 watts, 
then forced convection must be used.
WARNING
Install the instrument according to the enclosure protection 
provided. This instrument does not protect against the ingress of 
water. This instrument protects against finger access to hazardous 
parts within the enclosure.