Making Measurements
Using Markers
To Use Smith Chart Markers 
For greater accuracy when using markers in the Smith chart format, activate the discrete 
marker mode. Press 
 
 
.
To use Smith chart format:
1. Press 
 .
2. Press 
 
 
 and turn the front 
panel knob, or enter a value from the front panel keypad to read the resistive and 
reactive components of the complex impedance at any point along the trace. This is the 
default Smith chart marker.
The marker annotation tells that the complex impedance is capacitive in the bottom 
half of the Smith chart display and is inductive in the top half of the display.
• Choose 
 if you want the analyzer to show the linear magnitude and the 
phase of the reflection coefficient at the marker.
• Choose 
 if you want the analyzer to show the logarithmic magnitude and 
the phase of the reflection coefficient at the active marker. This is useful as a fast 
method of obtaining a reading of the log magnitude value without changing to log 
magnitude format.
• Choose 
 if you want the analyzer to show the values of the reflection 
coefficient at the marker as a real and imaginary pair.
• Choose 
 to show the real and imaginary parts of the device impedance 
(the series resistance and reactance, in ohms) at the marker. Also shown is the 
equivalent series inductance or capacitance.
• Choose 
 to show the complex admittance values of the active marker in 
rectangular form. The active marker values are displayed in terms of conductance 
(in Siemens), susceptance, and equivalent parallel circuit capacitance or inductance. 
Siemens are the international unit of admittance and are equivalent to mhos (the 
inverse of ohms).
Marker Fctn
 MKR MODE MENU   MARKERS:DISCRETE 
Format
 SMITH CHART 
Marker Fctn
 MARKER MODE MENU   SMITH MKR MENU 
 LIN MKR 
 LOG MKR 
 Re/Im MKR 
 R+jX MKR 
 G+jB MKR