1-38
 
Making Measurements
Using Markers
1. Press  .
2. Press 
 and turn the front panel knob, or enter a value from the front panel 
keypad to position the marker at a point of interest.
3. Press 
 to automatically add or subtract enough line length to the 
receiver input to compensate for the phase slope at the active marker position. This 
effectively flattens the phase trace around the active marker. You can use this to 
measure the electrical length or deviation from linear phase.
Additional electrical delay adjustments are required on devices without constant group 
delay over the measured frequency span.
Figure 1-27 Example of Setting the Electrical Delay Using a Marker 
Setting the CW Frequency 
1. To place a marker at the desired CW frequency, press:
 and either turn the front panel knob or enter the value, followed by a unit 
terminator.
2. Press 
  
.
You can use this function to set the marker to a gain peak in an amplifier. After pressing 
, activate a CW frequency power sweep to look at the gain 
compression with increasing input power.
Format
 PHASE 
Marker Fctn
 MARKER
DELAY 
Marker
Seq
 SPECIAL FUNCTIONS   MKR
CW 
 MKR
CW FREQ