Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
1-42
 
Making Measurements
Using Markers
To Calculate the Statistics of the Measurement Data 
This function calculates the mean, standard deviation, and peak-to-peak values of the 
section of the displayed trace between the active marker and the delta reference. If there is 
no delta reference, the analyzer calculates the statistics for the entire trace.
1. Move marker 1 to any point that you want to reference:
• Turn the front panel knob.
OR
• Enter the frequency value on the numeric keypad.
2. Press 
 
 
 to make marker 1 a reference marker.
3. Press 
 and move marker 2 to any position that you want to measure in 
reference to marker 1.
4. Press 
 
 
 to calculate and view the 
mean, standard deviation, and peak-to-peak values of the section of the measurement 
data between the active marker and the delta reference marker.
An application for this feature is to find the peak-to-peak value of passband ripple 
without searching separately for the maximum and minimum values.
If you are viewing a measurement in the polar or Smith Chart format, the analyzer 
calculates the statistics using the first value of the complex pair (magnitude, real part, 
resistance, or conductance).
Figure 1-32 Example Statistics of Measurement Data 
Marker
 
 MODE MENU   
 REF=1 
 MARKER 2 
Marker Fctn
 MKR MODE MENU   MKR STATS ON