Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
1- 77
Making Measurements
Using Limit Lines to Test a Device
Figure 1-61 Sloping Limit Lines 
Creating Single Point Limits 
In this example procedure, the following limits are set:
• from 
23 dB to 
28.5 dB at 141 MHz
• from 
23 dB to 
28.5 dB at 126.5 MHz
1. To access the limits menu and activate the limit lines, press:
 
 
 
 
 
 
2. To designate a single point limit line, as shown in 
, you must define two 
pointers:
• downward pointing, indicating the upper test limit
• upward pointing, indicating the lower test limit
Press:
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
System
 LIMIT MENU   LIMIT LINE
 LIMIT LINE ON   EDIT LIMIT LINE 
 CLEAR LIST   YES 
 ADD   STIMULUS VALUE 
141
M/
µ
 UPPER LIMIT 
23
x1
 LOWER LIMIT 
28.5
x1
 DONE 
 LIMIT TYPE   SINGLE POINT   RETURN 
 ADD   STIMULUS VALUE 
126.5
M/
µ
 UPPER LIMIT 
23
x1
 LOWER LIMIT 
28.5
x1
 DONE 
 LIMIT TYPE   SINGLE POINT   RETURN