HP (Hewlett-Packard) 6205B ユーザーズマニュアル

ページ / 68
5-47 To check the output stability, proceed as
a.
Connect test setup shown 
b. Turn CURRENT controls 
 clockwise
and adjust VOLTAGE controls for 40Vdc output.
c .
Allow 30 minutes warm-up then record
the differential voltmeter indication.
d. 
 8 hours, differential voltmeter
 change by less than 
 from indication
recorded in Step c.
5-48 TROUBLESHOOTING
5-49 Before attempting to troubleshoot this in-
strument, ensure that the 
 is with the instru-
ment and not with an associated circuit. The per-
formance test (Paragraph 5-10) enables this to be
determined without having to remove the instru-
ment from the cabinet.
5-50 A 
 understanding of the principles of
operation is a helpful aid 
 troubleshooting, and
 before attempting to troubleshoot
the unit in detail. Once the principles of opera-
tion are understood, refer to the overall trouble-
the symptom and probable cause.
5-51 The schematic diagram at the rear of the
manual (Figure 7-1) contains normal voltage 
TM 11-6625-2965-14&P
ings taken at various points within the circuits.
These voltages are positioned adjacent to the ap-
plicable test points (identified by encircled num-
bers). Component and test point designations are
marked directly on the main printed wiring board.
5-52 
 a defective component is located, replace
it and re-conduct the performance test. When a
component is replaced, refer to the repair and re-
placements and adjustment and calibration para-
graphs in this section.
5-53 OVERALL TROUBLESHOOTING PROCEDURE
5-54 To locate the cause of trouble 
 Steps
1, 2, and 3 in sequence.
(1) Check for obvious troubles such as
open fuse, defective power cord, input power fail-
ure, or defective voltage or current meter. Next
remove the top cover (held by four retaining
screws) and inspect for open connections, char-
red components, etc.
If the trouble source cannot
be detected by visual inspection, proceed with
Step 2.
(2) In almost 
 cases, the trouble can be
caused by improper dc bias or reference voltages;
thus, it is a good practice to check voltages in
Table 5-2, before proceeding with Step 3.
(3) Disconnect the 
Table 5-2.
Reference Circuit Troubleshooting
METER
METER
NORMAL
NORMAL
STEP
COMMON
POSITIVE
INDICATION
RIPPLE
IF INDICATION ABNORMAL, TAKE THIS ACTION
(P-P)
1
+s
33
6.2   0.3Vdc
Check 12.4 Volt bias or VR1 (See next paragraph)
2
31
+s
6.2   
Check 12.4 Volt bias or VR2 (See next paragraph)
3
+s
37
12.4 *1. 
Check Q8, Q9, 
 CR23, C1O, 
4
38
41
7.5   .7Vdc
Check C12, CR8, CR24, CR25
Table 5-3.
Overall Trouble shooting
SYMPTOM
CHECKS AND PROBABLE CAUSES
High output voltage
a.
Front 
 meter defective.
b. Series regulator feedback loop defective. Refer to 
 output voltage
a .
Fuses blown (Check 
 or C14 for short).
5-9