ASCII Fxxxx.x<CR> Sets the concentration factor to xxxx.x. For example: F1000<CR> will set the
factor to 1000.
Note <CR> is an ASCII carriage return character.
The last three commands provide an output which can readily be incorporated into most spreadsheet
software packages.
6.2 RS232 Output
The bi-directional RS232 interface is available on the rear panel 25 way D type connector.
The connections are as follows:
TXD 2
- INPUT TO 6300
RXD 3
- OUTPUT FROM 6300
RTS 4
- LINKED TO CTS
CTS 5
- LINKED TO RTS
DSR 6
- OUTPUT FROM 6300
DCD 8
- OUTPUT FRM 6300
DTR 20
- INPUT TO 6300 (must be active)
GND 7
Suggested interconnections are detailed below:
6300
IBM PC XT (25 way “D”)
TXD 2
2
TXD (From PC)
RXD 3
3
RXD (To PC)
RTS 4
4
RTS (From PC)
CTS 5
5
CTS (To PC)
DSR 6
6
DSR (To PC)
DCD 8
8
DCD (To PC)
DTR 20
20
DTR (From PC)
GND 7
7
GND
6300
IBM PC XT (9 way “D”)
TXD 2
3
TXD (From PC)
RXD 3
2
RXD (To PC)
RTS 4
7
RTS (From PC)
CTS 5
8
CTS (To PC)
DSR 6
6
DSR (To PC)
DCD 8
1
DCD (To PC)
DTR 20
4
DTR (From PC)
GND 7
5
GND
NOTE:
The Interface Cable Kit (Order Code: 542 009) can be used to implement the
above interconnections.
Recorder Output
This is available via the 4mm rear panel sockets. The level is proportional to the displayed reading, de-
pending on the measurement mode:
Transmission
1mV per 0.1%T
Absorbance
1mV per 0.001ABS
Concentration
1mV per concentration unit
13