the torque arm of the dynamometer to insure that there is no force 
being applied to the load cell (which would result in an inaccurate zero 
calibration).  Remove the calibration access cover panel and set the 
calibration mode switch to the ZERO position.  Also, be sure that 
the shunt resistor switch is set to the OFF position. At this time note 
that the RPM display of the unit should indicate “1111” and the 
POWER display should indicate “0000”.  The TORQUE display 
indicates the TORQUE and should read near zero. If the torque 
indication is excessive (plus or minus), the POWER display will read 
“EEEE” – which indicates that the zero offset of the load cell is 
excessive and that the ZERO calibration cannot proceed (see 
calibration notes below).  Otherwise, press and release the ENTER 
push-button switch to perform the zero calibration. At this point, the 
TORQUE display should indicate zero and the zero calibration 
procedure has been completed. 
2.)  Set the TORQUE SPAN calibration:  Apply the CALIBRATION 
TORQUE (see calibration torque notes below) to the unit and set the 
calibration mode switch to the SPAN position.  At this point the 
RPM display should indicate “2222”, the POWER display should 
indicate “0000”, and the TORQUE display should indicate a value 
approximately equal to the CALIBRATION TORQUE.  If the 
deviation between the torque reading and the calibration torque is 
excessive, the POWER display will read “EEEE” – indicating that the 
TORQUE SPAN calibration cannot proceed (see calibration notes 
below).  Otherwise, press and release the ENTER push-button switch 
to perform the TORQUE SPAN calibration. Now the TORQUE 
display should indicate the value of the CALIBRATION TORQUE 
and the torque span calibration has been completed. 
3.) 
Resume normal operation:  Set the calibration mode switch back to 
the OFF (center) position.  This turns off calibration mode and 
restores the unit to normal operation. Be sure to return the shunt 
resistor
 switch to the OFF position (if you used it during the torque 
span calibration). Replace the calibration access cover panel.  The unit 
has now been accurately calibrated.
 
 
Calibration Procedure Notes: 
 
1.  ZERO CALIBRATION cannot proceed.  This problem occurs when the ZERO 
OFFSET is excessive – too large for the zero offset correction facility of the 2400S to 
correct.  Check the following: 
1.1  Make sure that the load cell is properly connected to the 2400S horsepower 
computer.  Inspect the cable for damage, especially at the load cell end and the 
connector end.  Make sure the connector is securely fastened to the 2400S torque 
sensor connector. 
1.2  Make sure that the load has been completely removed from the load cell.  “Rock” 
the drive shaft of the dynamometer CW and CCW slightly (applying a small 
 
12