HP (Hewlett-Packard) E42 ユーザーズマニュアル
Model 5328A
Maintenance
Maintenance
Check
Point
Test
Observe on 5328A
1
REMOTE
Front panel (RMT) annunciator should be on.
2
CHECK
Counter should read 10.000 MHz.
3
RESOLUTION
The 9825A should print and counter display the 10 MHz
check signal with resolutions from 0.1 Hz to 1.0 MHz.
check signal with resolutions from 0.1 Hz to 1.0 MHz.
4
FREQ A
Counter display should read 0.0000 KHz.
5
RATlO B/A
Counter display should read 0.0000000
6
PERIOD A
Counter display should read 0. s
7
PER. AVE. A
Counter display should read 0.00000 ns
8
T.I. A-B
Counter display should read 0. s
9
T.I. AVG. A-B
Counter display should read 0.00000 ns
10
FREQ C
Counter display should read 0.0000 KHz
11
RATIO C/A
Counter display should read 0.0000000
12
SAMPLE RATE
When calculator displays MANUAL OK?, verify that front
SINGLE/
panel SAMPLE RATE control can be manually adjusted as
MULTIPLE
seen from GATE LIGHT flashing rate.
MEASMNT
When calculator displays GATE LIGHT OFF?, verify that Gate
Light is truly off.
Light is truly off.
5-42. Connect a function generator to the 5328 input channels and monitor the signal with an
oscilloscope as shown in Figure 5-8.
oscilloscope as shown in Figure 5-8.