JDS Uniphase RM3750 ユーザーズマニュアル

ページ / 41
RM3 SERIES BACKREFLECTION METER
USER’S MANUAL
 
10112341 Rev 002
Page 20 of 39
invoice number, and an itemized statement of claimed defects must be included with the return
material.
Ship return material in the original shipping container and packing material. If these are not
available, packaging guidelines are as follows:
 
1. Cover 
the 
front 
panel 
with foam to prevent damage.
 
2.  Wrap the unit in anti-static packaging. Use anti-static connector covers.
 
3.  Pack the unit in a reliable shipping container.
 
4. Use enough shock-absorbing material (10 to 15 cm) to cushion the unit and prevent it from
moving inside the container. Pink poly anti-static foam is the best material.
 
5.  Seal the shipping container securely.
6.  Clearly mark FRAGILE on its surface.
 
7.  Always provide the model and serial number of the unit and, if necessary, the RMA number
on any accompanying documentation.
8. In 
order 
to 
ship the unit please contact the JDS RMA department at 613-727-1303.
Cleaning Connectors
Caution
  Connecting damaged or dirty fibers to the unit can damage the connectors
on the unit.
  Never force an optical connector. Some connectors have a ceramic ferrule
that can easily be broken.
Optical cable ends need to be cleaned before using them with the unit.
The following items are required for cleaning:
  Filtered compressed air or dusting gas (for example, Tech Spray Envi-Ro-Tech Duster
1671 gas, available from http://www.techspray.com)
 Lint-free 
pipe 
cleaners 
(for example, from 3M
1
) or lint-free swab
  Lint-free towels (for example, 10 x 10 cm or 4 x 4 in HydroSorb III wipers, available from
http://www.focenter.com)
  Optical grade isopropyl alcohol or optical grade 200° ethanol (do not use rubbing alcohol,
which contains 30% water)
To clean the connectors:
                                               
1
 3M is a trademark of 3M.