Omega CNiS16 ユーザーズマニュアル

ページ / 72
Determine 
IN!1  and IN!2  Input Range and Resolution. For our transducer select
0 ~ 100 mV range and LOW resolution (10 µV)
IN!1  = 0 (mV) X 100 (cts/mV) x 1.0 = 0
IN!2  = 30 (mV) X 100 (cts/mV) x 1.0 = 3000
2. Determine correct values for Display Reading (
RD!1 and RD!2). In most cases,
RD!1 and RD!2 are equal to the minimum and the maximum of the transducer
output range.
RD!1 = 0000
RD!2 = 100.0
3. Scaling the controller.
Press 
d
28) Press 
d at the IN.RD prompt. Display shows IN!1  Input 1
Submenu.
Press 
d
29) Display shows last stored Input 1 value with 1
st
digit flashing.
Press 
b & c 30) Use b and c buttons to enter IN!1  value (0000).
Press 
d
31) Display advances to 
RD!1 only, if a change was made,
otherwise press 
a to advance to RD!1 Reading 1 Submenu.
Press 
d
32) Display shows last stored Reading 1 value with 1
st
digit
flashing.
Press 
b & c 33) Use b and c buttons to enter RD!1 value (0000). 
Press 
d
34) Display 
IN!2  Input 2 Submenu.
Press 
d
35) Display shows last stored Input 2 value with 1
st
digit flashing.
Press 
b & c 36) Use b and c buttons to enter IN!2  value (3000).
Press 
d
37) Display advances to 
RD!2 only, if a change was made,
otherwise press 
a to advance to RD!2 Reading 2 Submenu.
Press 
d
38) Display shows last stored Reading 2 value with 1
st
digit
flashing.
Press 
b & c 39) Use b and c buttons to enter RD!2 value (1000). 
Press 
d
40) Display flashes 
STRD stored message momentarily and
then advances to 
ALR1 only, if a change was made, otherwise
advances to 
ALR1 Alarm 1 Menu.
This scaling method based on 2 input values entered with 2
corresponding reading. Up to 10 linearization points can be
selected to customize the transducer curve. To select
linearization points see “L.PNt” Submenu.
24