Menu functions
Leica DISTO™ D5 LCA782204a en
EN
F
E
P
)
We recommend that measurements are initiated using the 
 
key when using the device on a tripod in order to prevent loss of sharp-
ness.
)
The settings are reset when the device is switched off.
Beep (
)
You can switch the beep on or off.
Offset (
)
An offset adds or subtracts a specified value automatically to or from all 
measurements. This function allows tolerances to be taken into account 
(e.g. unfinished dimensions compared with finished dimensions). If you 
selected the Offset function in the menu, you can now adjust the value 
using the 
 or the 
 key. Pressing the key for longer increases the 
rate of change of the values. When you reach the desired offset value, 
confirm it with the 
 key. The display shows the appropriate symbol 
 or 
 for as long as the offset value is set.
Digital pointfinder image black / white (
)
The display in pointfinder mode can be changed to black / white.
Level in status field (
)
The Level (in °) in status field can be switched on or off.
Reset - returning the instrument to the factory 
settings (
)
The instrument has a Reset function. If you select the menu function Reset 
and confirm, the device returns to the factory settings and stack and 
memory are deleted.
)
All customised settings and stored values are also lost.
Calibrate the tilt sensor (
)
You can calibrate the tilt sensor in the device. Calibration requires two 
measurements on a level surface.
Select calibration mode in the menu 
.
1
Carry out a first measurement 
 on a level surface. The device 
confirms the measurement with 
.
2
Rotate the device horizontally through 180° 
.
3
Press the 
 key and confirm that the device was rotated through 
180°.
4
Press the 
 key and take the second measurement. The device 
confirms the measurement with 
.
The tilt sensor is calibrated.