Agilent Technologies E4445A ユーザーズマニュアル

ページ / 664
166
Chapter 2
Instrument Functions: A - L
FREQUENCY / Channel
In
st
ru
m
e
nt
Funct
ions
:
A
-
L
Range:
Remote Command:
[:SENSe]:FREQuency:STOP <frequency>
[:SENSe]:FREQuency:STOP?
Example:
FREQ:STOP 1600
FREQ:STOP?
2.6.4
CF Step
Changes the step size for the center frequency and start/stop frequency functions. Once a
step size has been selected and the center frequency function is activated, the step keys
(and the UP|DOWN parameters for Center Frequency from remote commands) change
center frequency by the step-size value. The step size function is useful for finding
harmonics and sidebands beyond the current frequency span of the analyzer. When
auto-coupled in a non-zero span, the center frequency step size is set to 10% of the span.
Key Path:
FREQUENCY
State Saved:
Saved in Instrument State
Factory Preset: Span/10
Knob
Increment:
Step Key
Increment:
Model
Frequency Range
E4440A
–100.0000000 MHz
to
27.0 GHz
E4443A
–100.0000000 MHz
to
13.7 GHz
E4445A
–100.0000000 MHz
to
7.2 GHz
E4446A
–100.0000000 MHz
to
44.5 GHz
E4448A
–100.0000000 MHz
to
51.0 GHz
if
Span = 0 Hz
increment = RBW/20
if
Span > 0 Hz
increment = Span/200
if
CF Step = auto, Span = 0 Hz
step = RBW
if
CF Step = auto, Span > 0 Hz
step = Span/10
if
CF Step = manual
step = 1, 2, 5 sequence