HP (Hewlett-Packard) 8360 ユーザーズマニュアル

ページ / 508
Function Group
MARKER
Menu Map 
Description
This 
 sets the center frequency of the sweep to the frequency
of the most recently activated marker. Select any marker Ml . . .
 
then select 
 to change the center frequency of
the sweep to that of the marker. The frequency span does not change
unless the new sweep limits fall outside the frequency range of the
synthesizer, in that case the synthesizer automatically scales the
frequency span to be within the synthesizer’s operating frequency
range.
Programming Codes
 ?
FREQuency:CENTer 
 from above> [freq suffix]
Analyzer: 
MC
See Also
“Marker Operation” in Chapter 1.
Clear Fault
Function Group
Menu Map
Description
Programming Codes
See Also
SERVICE
6
This 
 clears all the latched fault status indicators.
SCPI: 
DIAGnostics:OUTPut:FAULts
The above command relays the fault information and clears all faults.
Analyzer: 
NONE
Fault 
Menu
 Operating and Programming Reference
HP 8360
User’s Handbook