HP (Hewlett-Packard) 8360 ユーザーズマニュアル
Coupling Factor
Function Group
Menu Map
Description
Programming Codes
See Also
ALC
1
This
allows specification of the coupling factor of an external
coupler/detector used to externally level the synthesizer output
power. Negative coupling factor values are required for valid entry.
See “Specifications” for the coupling factor range.
power. Negative coupling factor values are required for valid entry.
See “Specifications” for the coupling factor range.
SCPI:
POWer:ALC:CFACtor
Analyzer: NONE
A
ALC
“Externally Leveling the Synthesizer” in Chapter 1.
C
CW
Function Group
Menu Map
Description
Programming Codes
See Also
FREQUENCY
2
This
lets you select a synthesized continuous wave frequency.
When you press
the synthesizer stops sweeping (green SWEEP
LED off) and displays:
CW: XXXXX
MHz. Where XXXXX
represents a frequency value. Use either the rotary knob, the step
keys (with or without the left/right arrow keys), or the numerical
entry keys with a terminator key to set the desired value. If a small
change is desired, use the left and right arrow keys to underline the
digit to be changed.
keys (with or without the left/right arrow keys), or the numerical
entry keys with a terminator key to set the desired value. If a small
change is desired, use the left and right arrow keys to underline the
digit to be changed.
FREQuency[:CW] <num>[freq suffix] or
FREQuency:MODE CW
Analyzer:
CW
Coupled,
“CW Operation and Start/Stop Frequency Sweep” in Chapter 1.
“Programming Typical Measurements” in Chapter 1.
“Programming Typical Measurements” in Chapter 1.
Operating and Programming Reference HP 8360
User’s Handbook