M
 Sweep
Function Group
MARKER
Menu Map 
Description
This 
 lets you set the synthesizer to start sweeping at the
frequency of marker 1 (Ml), and stop sweeping at the frequency of
marker 2 
 
 must have a higher frequency value than Ml. If
 Sweep 
is activated when 
 is at a lower frequency than
Ml, the values of Ml and 
 are permanently interchanged. While
this function is active, the start/stop frequencies of the synthesizer
are changed to the values of Ml and 
 An asterisk next to the key
label indicates this feature is active.
Programming Codes
SCPI: 
SWEep:MARKer:STATe 
Analyzer: 
 function on, MPO function off.
See Also
Marker Ml, 
 
“Marker Operation,” in Chapter 1.
“Programming Typical Measurements,” in Chapter 1.
Manual Sweep
Function Group
SWEEP
Menu Map 
Description
This 
 lets you set the synthesizer to the manual sweep mode
of operation. Depending on what parameter is sweeping, you can
use either the rotary knob or the ARROW keys to manually sweep
between the start/stop limits. In manual sweep mode, the synthesizer
does not automatically retrace at the sweep end point (the user must
turn the rotary knob to retrace), and the green SWEEP LED does
not light. The resolution of the rotary knob is 0.1% of the sweep
span in either start/stop or 
 mode. The resolution of the
 and 
 arrow keys are dependent on the resolution defined by
HP 8360
User’s Handbook
Operating and Programming Reference M-l