HP (Hewlett-Packard) 8360 ユーザーズマニュアル

ページ / 508
Power Offset
Function Group
POWER
Menu Map 
Description
This 
 changes the mapping of absolute power parameters on
input to the synthesizer. It does not change the RF output produced
by the synthesizer. The equation used to determine the displayed
value is:
Entered or Displayed Power = Hardware Power (ALC)   Active
Offset.
Programming Codes
POWer:OFFset:STATe 
POWer:OFFset 
Analyzer: 
NONE
 
[
POWER LEVEL
and POWER 
Power Slope
Function Group
POWER
Menu Map 
Description
This 
 lets you compensate for system, cable, and waveguide
variations due to changes in frequency, by linearly increasing or
decreasing power output as the frequency increases. RF slope values
may range from -2.50 to 
 
 per 
 The power at the
beginning of the sweep equals the current power level. An asterisk
next to the key label indicates that this feature is active.
Programming Codes
POWer:SLOPe:STATe 
POWer:SLOPe 
Analyzer: 
 function on, SLO function off. Note that because SL
functions in the fundamental units of 
 you program the SL
code 
 where m is 1 (on) or 0 (off); d is the numerical value
 Operating and Programming Reference
HP 8360
User’s Handbook