HP (Hewlett-Packard) 8360 ユーザーズマニュアル

ページ / 508
General
Environmental
Warm-Up Time
Power Requirements
Weight   Dimensions
Adapters Supplied
Operating Temperature Range: 0 to 
EMC: Within limits of VDE 
 Level B, FTZ 
 and
 Part 7 
Operation: Requires 30 minute warm-up from cold start at 0 to
 Internal temperature equilibrium reached over 2 hour warm-up
at stable ambient temperature.
Frequency Reference: Reference time base is kept at operating
temperature with the instrument connected to AC power.
Instruments disconnected from AC power for more than 24 hours
require 30 days to achieve time base aging specification. Instruments
disconnected from AC power for less than 24 hours require 24 hours
to achieve time base aging specification.
48 to 66 Hz; 115 volts 
 or 230 volts 
 400
VA maximum (30 VA in standby)
Net Weight: 27 kg (60 lb)
Shipping Weight: 36 kg (80 lb)
Dimensions: 178 H x 425 W x 648 mm D (7.0 x 16.75 x 25.5 inches)
HP 
 
 
 
 
Type-N (female)   3.5 mm (female)
Part number 1250-1745
3.5 mm (female)   3.5 mm (female)
Part number 5061-5311
HP 
 
2.4 mm (female)   2.92 (female)
Part number 1250-2187
2.4 mm (female)   2.4 mm (female)
Part number 1250-2188