 100 MHz VCXO, Troubleshooting
The 100 MHz VCXO supplies all of the fixed reference frequencies for the HP 8566B. Separate
100 MHz outputs are routed to the 
 
 and 
 assemblies. A 400 MHz signal is
applied to the 
 assembly. The 
 assembly divides the 100 MHz input to provide 10
MHz and 20 MHz references.
100 MHz Oscillator 
The tune voltage at 
 is nominally -8 V. If an unlocked condition exists, this voltage
is either -23 V or -1.3 V. If the level is -23 V, check the 10 MHz reference to 
 If the
level is -1.3 V, check the outputs of the 100 MHz Buffer/Amplifier.
Typical voltage levels for the oscillator transistor Q5 are:
Emitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Collector
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
..O.O V
100 MHz Buffer/Amplifier 
All 100 MHz output levels should be approximately 0 
 Typical voltage levels for each
transistor are as shown in Table 1.
 
 
Emitter
4.0 3.9
3.9
Base
4.7 4.7 4.6
Collector 9.3 9.4
9.4
Table 1. 100 MHz Buffer/Amplifier Voltage Levels
Quadrupler   and 400 MHz Amplifier 
Typical voltage levels for the Quadrupler and 400 MHz Amplifier are as shown in Table 2.
The 400 MHz Amplifier output at 
 (96 cable) is approximately -10 
 when the
 100 MHz board assembly is in the casting.
Table 2. Quadrupler and 400 MHz Amplifier Voltage Levels
Q3 Q4 
 
Q3
Emitter
-4.6 -5.1 -6.6 -6.6
Base
-5.9 -5.9 -5.9 -5.9
Collector
0.0
0.0
0.0
0.0