HP (Hewlett-Packard) HP 85660B ユーザーズマニュアル
The second and third numbers are the M and N divide numbers respectively. The waveform
at test points 3 and 7, 5 and 6, and 1 and 2 should be as shown in Figures 1, 2, and 3
respectively. The Center Frequencies for Figure la, lb, lc, and
at test points 3 and 7, 5 and 6, and 1 and 2 should be as shown in Figures 1, 2, and 3
respectively. The Center Frequencies for Figure la, lb, lc, and
are 3.77
3.96
4.15
and 4.34
respectively. The M and N numbers are 20, 21, 22, and 23
respectively.
Oscilloscope control settings for Figure 1, Figure 2, and Figure 3 are:
Vertical: 0.1
Horizontal: 0.2
Probe:
Horizontal: 0.2
Probe:
Note
Waveforms on the M Divider require the M/N VCO be phase-locked at 380
MHz. If the M/N VCO is not phase-locked, an external signal generator may
be used to locate the failure.
MHz. If the M/N VCO is not phase-locked, an external signal generator may
be used to locate the failure.
Note
3
All waveform voltage levels are emitter coupled logic (ECL) levels:
-0.9 V high, -1.7 V low.
Mixer and IF Amplifier
The Mixer inputs are 400 MHz from
100 MHz VCXO and 355 to 395 MHz from
M/N Output. When the M and N numbers are the same, the M/N output frequency is
380 MHz.
380 MHz.
3