The second and third numbers are the M and N divide numbers respectively. The waveform
at test points 3 and 7, 5 and 6, and 1 and 2 should be as shown in Figures 1, 2, and 3
respectively. The Center Frequencies for Figure la, lb, lc, and 
 are 3.77 
 3.96
 4.15 
 and 4.34 
 respectively. The M and N numbers are 20, 21, 22, and 23
respectively.
Oscilloscope control settings for Figure 1, Figure 2, and Figure 3 are:
Vertical: 0.1 
Horizontal: 0.2 
Probe: 
Note
Waveforms on the M Divider   require the M/N VCO be phase-locked at 380
MHz. If the M/N VCO is not phase-locked, an external signal generator may
be used to locate the failure.
Note
3
All waveform voltage levels are emitter coupled logic (ECL) levels:
-0.9 V   high, -1.7 V   low.
Mixer   and IF Amplifier 
The Mixer inputs are 400 MHz from 
 100 MHz VCXO and 355 to 395 MHz from 
M/N Output. When the M and N numbers are the same, the M/N output frequency is
380 MHz.
 3