 Trace A 
 
 displays negative peaks.
(SHIFT) Trace A 
 
 displays sampled data.
(
S
  
 
 
 
 
 
 
H
I
F
T
)
 
T
r
a
c
e
 
T
 
T r a c e   A ( k s a ) r e t u r n s   t o   t h e   n o r m a l   a u t o m a t i c   d e t e c t i o n   m o d e s   a n d
removes the CRT indication.
Error Correction Routine
The internal Error Correction Routine available by pressing 
 
 
(KSW) is also useful as a diagnostic aid. If a malfunction causes the routine to stop, restart
it and note the control settings (RES BW, ATTEN,  REF LEVEL, LOG/LIN,  etc.) when
the failure occurred. If the routine runs, the correction factors can be displayed by pressing
 
 
 Figure 2 displays the data for a typical instrument. Table 2 gives the
parametric information, specifications and a place to start the troubleshooting procedure.
Caution must be exercised in interpreting the correction factor data. Wrong conclusions
can be reached by not understanding how the internal program runs. The program assumes
that the input signal level is -10.0 dBm. Any error in this level translates to the correction
factors; therefore, the Amplitude Accuracy test 
 8 should be performed first. The
internal program runs in the LIN mode while 
 8 is in 1 dB/LOG mode. Large offsets
in LOG/LIN offset (lines 1 and 14) will cause errors in the data.
  - 7 . 0  
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 2. Error Correction Routine Data
8 Overall Troubleshooting