HP (Hewlett-Packard) HP 85660B ユーザーズマニュアル
Coaxial Switch and
300 MHz BPF Replacement
Note
For location of hardware and cables referred to in this procedure, see Figure 2.
1. Remove front panel from RF Section. (See A5 Front Panel Section for front panel removal
procedure.)
2.
Replacement
a. Slightly loosen connectors
and
b. Unsolder wires from the coaxial switch.
1 (brown) wire from input
3 (orange) wire from input
2 (red) wire from input
c. Release coaxial switch by removing two screws
d. Remove coaxial switch from three cables by completely disconnecting at connectors
and
e. Resolder three wires to new coaxial switch.
f. Connect three cables to coaxial switch
and
Do not tighten connectors.
g. Attach new coaxial switch to instrument with two screws
h. Tighten connectors to coaxial switch.
3.
Replacement
a. Disconnect connectors and and remove
BPF from mounting clips.
b. Install new
BPF in mounting clips and reconnect connectors.
4. Replace front panel. (See A5 Front Panel Section for front panel replacement procedure.)
A6 3