HP (Hewlett-Packard) HP 85660B ユーザーズマニュアル

ページ / 570
 Second Converter Replacement
Note
For location of hardware and cables referred to in this procedure, see Figure 4.
Removal
1. Remove ac line cords and RF Section bottom cover.
2. Disconnect connectors 
   and 
 single-pin connector 
 and double-pin connector 
from Second Converter.
3. Disconnect the remaining cables from Second Converter
1 (brown) cable from 
 OUT 
4 (yellow) cable from TUNE 
4. Remove two screws   to remove Second Converter from A6 RF Module.
Installation
5. Position new Second Converter on mounting bracket. Secure with two screws 
6. Reconnect connector 
 connector 
 single-pin connector 
 and double-pin connector 
to Second Converter.
7. Reconnect the following cables to Second Converter.
1 (brown) cable to 
 OUT 
1 (brown) cable to 
 IN 
4 (yellow) cable to TUNE 
8. Replace bottom cover and ac line cords.
8 A6