 MICROLAB
®
  Carousel User Guide (June 2003)
 15
Calibration
Calibration sets an offset angle from the default initialization position.
To calibrate the carousel:
1. Select Start > Programs > Hamilton > Carousel > Service 
Software from the Windows desktop. The Service Software 
window appears. 
2. Select Communication > Connect. A Select Instrument window 
appears.
3. Select the Carousel and click OK. The File transfer icon (arrows) on 
the tool bar should turn from gray to blue if communications are 
established.
4. Click on Instruments Carousel > Calibration. The carousel will 
initialize, and a Calibrate Carousel window will appear (figure 14).
Figure 14:
Calibrate Carousel window
5. Use the left and right arrow keys to rotate the carousel. Each 
keystroke will move the platter the number of degrees shown in the 
window (refer to figure 14). Change that resolution as needed. 
Move the platter until position #1 aligns with the intended plate pickup 
position. In most cases this is 90
o
 counter-clockwise from the index 
pointer (figure 15)
Do not rotate the carousel platter by hand.