Philips V200 ユーザーズマニュアル

ページ / 328
Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-31
Chapter 13
2. Press "Monitor Setup" on a Standard Control Panel, or "Instrument 
Config." on a Classic Control Panel of the Component Monitoring 
System (CMS). 
3. Highlight "Operating Modes" on the displayed window and press 
"Confirm". The current operating mode will appear in the task window 
(for example, "Operating Mode - Monitoring") along with a prompt for 
the password.
4. Enter the password, 1245. If the password is entered correctly, the 
"Change OpMode" key will be highlighted.
5. Press "Change OpMode" and select "Config" or "Monitor".
6. Press "Confirm". 
a. If "Config" was selected in step 5, the display will become blank 
and the message "Config mode active - NO MONITORING!" will 
appear at the top of the screen. Continue with the "Preselecting 
Devices" section below.
b. If "Monitor" was selected in step 5, the VueLink system will begin 
to display data. Make sure that the LED next to "Open Device" is 
lit while in Monitoring mode to display ventilator data. If it is not 
lit, press the button on the VueLink Module, select "Open 
Interface" and press confirm. Pressing the button on the VueLink 
Module will then display the Standard Parameter Interface (SPI) 
screen. Pressing the VueLink Module button a second time will 
activate the ventilator screen with more information. 
Preselecting Devices
Once in Configuration Mode, the Open device must be selected as one of the 
three preselected devices for the VueLink Ventilator Module. To do this, follow 
the steps below. 
1. Press "Module Setup" or "Module Config" on the CMS.
2. Highlight "VENTILTR", "OPEN INTRFACE", OR "ESPRIT VENT" and 
press "Confirm".
3. Note: Only one of the above choices will be available at a time. The 
choice depends on how the module is already configured.
4. Verify "Prefer'd Module" is set to "Ventiltr" and choose whether device 
alarms should be Accepted or Ignored.
Press "Module Config".
5. Highlight "Open Interface M1032-TU1AA" by using the arrow keys or 
pressing "Next Device".
6. Press "On/Off Preselect" if "On" is not already displayed beside this 
item.