Philips V200 ユーザーズマニュアル

ページ / 328
Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-97
Chapter 13
Neonatal Option
The Neonatal option allows the Respironics V200 Ventilator to ventilate 
intubated neonatal patients with an ideal body weight range of 0.5 - 6.5 kg 
(1.10 - 14.33 lb.) and an endotracheal tube I.D. range from 2.5 - 4.0 mm. 
The option provides pressure control in A/C, SIMV, and Apnea ventilation and 
also provides pressure support in SIMV and CPAP. The Neonatal Option is not 
available in VCV or NPPV modes.
System Requirements
The Neonatal option requires the following software and accessories for proper 
use:
Figure 13-52: Circuit to Filter Connection
Neonatal option Requirements
Respironics V200 Ventilator
Neonatal option loaded and enabled.
NOTE:
Ventilators with the Neonatal option software 
installed are identified by the option label 
located near the power switch.
Recommended Accessories
Part Number
Neonatal patient circuits (10 mm I.D.)
Fisher & Paykel reusable circuit 
Hudson RCI disposable circuit
Fisher & Paykel disposable circuit 
22mm, Male x Male Connector (Refer to Figure 13-52.)
coupling, Straight Silicone
900MR780
780-07
RT131
(RT125 is EU equivalent)
1002505
500-1000-43
NOTE:
We recommend using approved circuits or equivalent when connecting 
to the ventilator.
Table 13-20: Neonatal Option Requirements
Filter
Connector
Circuit
Silicone coupler