Philips V200 ユーザーズマニュアル

ページ / 328
Appendix A
RS-232 Communications Protocol
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
Appendix A-21
Alarm Status Report 
(ALRM)
When the ventilator receives ALRM followed by a carriage return, it will 
respond by transmitting the information shown in Table A-6: “Alarm Status 
Report”. The ventilator responds to the ALRM command by returning a string 
with a variable length. Fields 2 through 4 define the length of the message. 
The last character transmitted is a stop code indicating the end of the 
message. The second field indicates the number of characters between the 
start and stop codes. The third field indicates the number of fields between 
the start and stop codes. The fourth field is the start code, 0x02. The last field 
in the string is the stop code, 0x03.
Static Compliance
19.2
◆◆
0.01 for 0.00 to 
9.990.1 for 10.0 to 
99.91 for 100 to 
350
0.00 - 350
mL/
cmH2
O
Static Compliance 
display from the 
Static C & R 
Screen
Static C & R Time 
of Last Maneuver
11-14-01
◆◆13:11
1 minute
01-01-70 
0:00 to
01-19-38
3:14
N/A
Static C & R Time 
of Last Maneuver 
display from Static 
C & R Screen.15 
character field, 24 
hour clock: MM-
DD-YYwwHH:MM
AutoPeep
1.2
◆◆
0.1 for -9.9 to 9.91 
for -10 to -201 for 
10 to 120
-20 - 120
cmH2
O
AutoPeep display 
from the 
Mechanics Patient 
Status Screen
AutoPeep Time of 
Last Maneuver
11-14-01
◆◆13:11
1 minute
01-01-70 
0:00 to
01-19-38
3:14
N/A
AutoPeep Time of 
Last Maneuver 
display from 
Mechanics Patient 
Status Screen.15 
character field, 24 
hour clock: MM-
DD-YYwwHH:MM
Static C & R 
Plateau End 
Pressure
8.2
◆◆◆
0.2 for -20 to 99.9
1 for 100 to 130
-20.0 - 130 cmH
2
O P
plat
 display from 
Static C & R 
Screen
Stop Code
0x03
N/A
N/A
N/A
ASCII End 
Transmission 
Character (ETX)
Patient Data Report (Continued)
Description
Example
Resolution
Range
Units
Comments
Table A-5: Patient Data Report (Sheet 4 of 4)