Chapter 2
Function Generator Operation
©
 National Instruments Corporation
2-11
Pre-Attenuation Offset
The NI 5401 hardware supports a DC offset of up to ±2.5 V before the 
attenuation chain. Unless the 10 dB attenuator is switched in, which occurs 
when the amplitude is less then 3.16 V
pk-pk
, the waveform maximum plus 
the offset must not exceed ±5 V into 50
Ω
. If it does, the waveform is 
clipped. Refer to Figure 2-8 for a diagram showing the location of the 
10 dB attenuator.
NI-FGEN automatically calculates the pre-attenuation offset value based 
on the DC offset and amplitude values, so the allowable DC offset range is 
dependent on the amplitude. For example, if you have an amplitude of 
1 V
pk-pk
, the maximum DC offset you can apply is 0.25 V, which 
corresponds to a pre-attenuation offset of 2.5 V.
Note
You can change the DC Offset at any time during waveform generation. Refer to 
your software documentation for additional information.
Phase-Locked Loops and Board Synchronization
Figure 2-11 illustrates the block diagram for the NI 5401 for PCI PLL 
circuit. Figure 2-12 illustrates the block diagram for the NI 5401 for PXI 
PLL circuit. The PLL consists of a voltage-controlled crystal oscillator 
(VCXO) with a tuning range of ±100 ppm. This VCXO generates the main 
clock of 80 MHz. 
The PLL can lock to a reference clock source from the external connector, 
from a RTSI Osc line on the RTSI bus (NI 5401 for PCI), or from a 
10 MHz Osc line on the PXI backplane bus (for NI 5401 for PXI). The PLL 
can also be tuned internally using a calibration DAC (CalDAC). National 
Instruments accurately performs this tuning during manufacturing. Refer to 
the 
 section later in this manual for additional 
information on using the RTSI and 10 MHz Osc lines.
The reference and VCXO clock are compared by a phase comparator 
running at 1 MHz. The loop filters the error signal and sends it to the control 
pin of the VCXO to complete the loop.